Steam Sterilizer (CS-ST045, ST065)

Product Detail Information

Heater Compartment
With a separate heater compartment at the bottom of the chamber, the sterilizer not only has a larger capacity than other products, but also utilizes less water for sterilization. The design minimizes heating time and greatly improves drying performance.

Air Evacuation after Water Supply
The method of vacuum drying after water is supplied to the chamber differs from existing processes in that all air in the chamber is removed completely with a vacuum pump before water is added, to assure 100% sterilization. The conventioanl water supply method allows a considerable volume of air on the objects of sterilization, so that a through and complete sterilization is not guaranteed. Certain spores in an air layer may not be completely removed. The Air Evaporation Method reduces the evaporation point of the water and forcibly releases all the air in the objects to be sterilized, guaranteering perfect sterilization results with quick heating and steam penetration.

Appealing Design
The products is beautifully designed for a refined appearance and harmonious integration in a hospital environment.

Quick Water Supply
By supplying water directly into a vacuum, the filling time is sharply reduced (20 seconds at most) and the product is usable continously at high temperatures. It supplies water faster than with conventional methods.

Fully automated Control
The whole process of sterilization, from vacuum sterilization to vacuum drying is electronically controlled and fully automatic. Once the sterilization program is set, only press START and the complete process of Vacuum Build-up → Water Supply → Heating → Sterilization → Air Release → Vacuum Drying takes place automatically.

Recorder Installation
The printer records sterilization conditions (date, temperature, time and drying time, etc.) under normal operation.

Self-diagnosis
A total of 13 error conditions are displayed under the self-diagnostic control. Your Safety is our top Priority!

This steam sterilizer is designed for user's convenience and patient's safety preferentially.

Self-diagnosis:

A total of 13 error conditions are displayed under the self-diagnostic control.
ERROR-1
Water tank low
ERROR-2
Heater overheated (Standard throughout the process : 155°C, Dry : 190°C
ERROR-3
Door open during operation
ERROR-4
Sterilization temperature deviations
ERROR-5
Short water supply
ERROR-6
Heating time exceeded
ERROR-7
Temperature sensor error(Disconnection/Short circuit)
ERROR-8
Heater temperature sensor error(Disconnection/Short circuit)
ERROR-9
Pressure sensor error(Disconnection/Short circuit)
ERROR-10
Vacuum release test failure
ERROR-11
Vacuum lead time excess setting
ERROR-12
Chamber overheated during water supply
ERROR-13
Chamber pressure exceeded (2.4kgf/cm2)

LCD Display :

On the control part, both the LCD display and FND display the steps of the sterilization process. The displays are large and easily visible, even from a distance. The LCD displays sterilzation temperature, pressure, lead time, drying temperature and time, heater compartment temperature, water tank level, and door open/close status, so that the user can monitor the equipment fully at any time.

Sterilization Programs :

The test program mode is the most convenient to use. With the Bowie-Dick test, a special program to check sterilizer performance on a regular basis, the efficiency and reliability of the equipment is maintained. The VACUUM LEAKAGE TEST shows any leaks in the sterilizer, so that the user can determine firsthand whether the product has a problem.

Basic program
 
Pack
Instrument
Glove
Liquid
Termperature
134°C
134°C
121°C
131°C
Pressure
2kgf/cm2
2kgf/cm2
1kgf/cm2
1kgf/cm2
Time
20min
10min
30min
30min
Dry Time
30min
20min
20min
-

User program
 
Manual program
Dry program
Termperature
121°C~134°C
-
Pressure
1kgf/cm2 ~ 2kgf/cm2
-
Time
0~60min
-
Dry Time
0~60min
0~60min
Test program
 
Vacuum leakage test
Bowie-dick test
Termperature
Normal temperature
134°C
Pressure
0kgf/cm2
2kgf/cm2
Time
15min
4min
Dry Time
-
4min

Specifications

Model
CHS-ST045
CHS-ST065
Overall Site
580(W) x 510(H) x 650(L)mm
580(W) x 510(H) x 820(L)mm
Chamber
Size
310(W) x 340(H) x 430(L)mm
310(W) x 340(H) x 620(L)mm
Type
Square
Material
KS D 3577-85(STS316)
Capacity
45Liter
65Liter
Sterilizer Temperature, Pressure
121°C(1kgf/cm2) ~ 134°C(1kgf/cm2)
Reservoir Capacity
6.0Liter
8.0Liter
Heater
2.3KW
2.9KW
Power Source
AC220V 50/60Hz 2.3W
AC220V 50/60Hz 2.9W
Air Removal
Oil-less vacuum pump(40 RNS)
Oil-less vacuum pump(50 RNS)
Display
Graphic LCD
Control Mode
Micro processor
Recorder
Dot Printer
Dry System
Vacuum dry system
Weight
93kg
110kg

Option Accessory

Cart
CHS-ST045
620(W) x 820(H) x 560(L)mm
CHS-ST065
620(W) x 820(H) x 750(L)mm
Printer
Dot Printer
Product Detail Image